![]() | Range of Tecan hybridization chambers for high application flexibility |
The HS Pro Hybridization Station offers several chamber formats for various microarray dimensions:

Chamber 51x20
| Dimension [mm] | ||||
| Min | Max | |||
| Slide | A | 25 | 26 | |
| B | 75.3 | 76.5 | ||
| C | 0.8 | 1.2 | ||
| Recommended area for spotting | Array | a | 3.2 | |
| b | 18.8 | |||
| c | 21.8 | |||
| d | 45.9 | |||
| Barcode | e | 15.6 | ||
Chamber 57x21
| Dimension [mm] | ||||
| Min | Max | |||
| Slide | A | 25 | 26 | |
| B | 75.3 | 76.5 | ||
| C | 0.8 | 1.2 | ||
| Recommended area for spotting | Array | a | 2.7 | |
| b | 19.8 | |||
| c | 15.8 | |||
| d | 51.9 | |||
| Barcode | e | 9.6 | ||
Chamber 63.5x21
| Dimension [mm] | ||||
| Min | Max | |||
| Slide | A | 25 | 26 | |
| B | 75.3 | 76.5 | ||
| C | 0.8 | 1.2 | ||
| Recommended area for spotting | Array | a | 2.7 | |
| b | 19.8 | |||
| c | 9.3 | |||
| d | 58.4 | |||
| Barcode | e | 3.0 | ||
Chamber A1x244k (61.5x21)
| Dimension [mm] | ||||
| Min | Max | |||
| Slide | A | 25 | 26 | |
| B | 75.3 | 76.5 | ||
| C | 0.8 | 1.2 | ||
| Recommended area for spotting | Array | a | 2.7 | |
| b | 19.8 | |||
| c | 13 | |||
| d | 57.9 | |||
| Barcode | e | 8.5 | ||

Segmented Chamber: Dual Active Area Chamber 27.8x20
| Dimension [mm] | ||||
| Min | Max | |||
| Slide | A | 25 | 26 | |
| B | 75.3 | 76.5 | ||
| C | 0.8 | 1.2 | ||
| Recommended area for spotting | Array | a | 3.2 | |
| b | 18.8 | |||
| c1 | 45.6 | |||
| c2 | 13.7 | |||
| d | 22.7 | |||
| Barcode | e | 7.6 | ||
| Barcode | e | 8.5 | ||
A2x105k Chamber: Dual Active Area Chamber 28.2x21
| Dimension [mm] | ||||
| Min | Max | |||
| Slide | A | 25 | 26 | |
| B | 75.3 | 76.5 | ||
| C | 0.8 | 1.2 | ||
| Recommended area for spotting | Array | a | 2.7 | |
| b | 19.8 | |||
| c1 | 45.4 | |||
| c2 | 13.7 | |||
| d | 24.6 | |||
| Barcode | e | 8.5 | ||
| Barcode | e | 8.5 | ||

QuadChamber A4x44k (22x13)
| Dimension [mm] | ||||
| Min | Max | |||
| Slide | A | 25 | 26 | |
| B | 75.3 | 76.5 | ||
| C | 0.9 | 1.1 | ||
| Recommended area for spotting | Array | a | 2.6 | |
| b | 19.8 | |||
| c1 | 12.6 | |||
| c2 | 28.3 | |||
| c3 | 44 | |||
| c4 | 59.8 | |||
| d | 11.8 | |||
| Barcode | e | 10.3 | ||
© 2008, Tecan Trading AG, Switzerland, all rights reserved. Disclaimer | Trademarks | Site Map