| Detection mode | Absorbance |
| Light source | LED, self-calibrating |
| Photodetector | silicon photodiodes |
| Plate type | 96-well microplates (including strip-well microplates) |
| Measurement modes | Endpoint, kinetic and multi-label measurements |
| Shaking | Linear shaking in 4 different modes |
| Measurement channels | 8 |
| Measurement range | 0 – 4 OD |
| Resolution | 0.0001 OD |
| Wavelength range | 400 – 750 nm |
| Wavelength selection | up to 8 filters on wheel |
| standard filters 405, 450, 492, 620 nm* | |
| Accuracy | |
| 0.000 – 2.000 OD | ≤ (0.5 % + 0.010 OD) typical |
| 2.000 – 3.000 OD | ≤ (1 % + 0.010 OD) typical |
| Precision / Reproducibility | |
| 0.000 – 2.000 OD | ≤ (0.5 % + 0.005 OD) |
| 2.000 – 3.000 OD | ≤ (1.0 % + 0.005 OD) |
| Linearity | |
| 0.000 – 2.000 OD | ≤ 1 % |
| 2.000 – 3.000 OD | ≤ 1.5 % |
| Computer interface | USB 1.1 / 2.0 |
*additional filters available separately