Skip to content
  • There are no suggestions because the search field is empty.

ERR009 LLD Error

1. Usual root causes

An issue causes the liquid level detection option to detect an empty liquid or full waste bottle.

2. First aid
• Check the filling level of the bottles:

  • Check whether the liquid bottle used for the selected input channel is filled with liquid.
  • Check whether the waste bottle is empty.

• Make sure that the liquid bottle floating switch wire that is used for the selected input channel is connected correctly to the back side of the instrument.
• Make sure that the waste bottle floating switch wire is connected correctly to the back side of the instrument.

3. Typical solution
If the issue cannot be resolved, contact Tecan® to schedule an onsite intervention or to send the instrument for in-house repair.

4. Further recommendations
This error can be avoided by ensuring proper connections between the attached bottles and the instrument prior to starting a program.

5. Next steps
• Note your instrument serial number.
• If the instrument is controlled by HydroControl™, save the log files.
• Create a service request and include the serial number and log files.

Helpdesk Tecan

 

Further Help

Hydroflex™_How To 

 

Link to Tecan Product Page

HydroFlex™ Plus microplate washer

 

401776-011