ERR031 Error LLD: Bottle Waste Full

1. Usual root causes

An issue causes the liquid level detection option to detect a full waste bottle.


2. First aid

• Check the filling level of the bottle:

  • Check whether the waste bottle is empty.

• Make sure that the waste bottle floating switch wire is connected correctly to the back side of the instrument.


3. Typical solution

If the issue cannot be resolved, contact Tecan® to schedule an onsite intervention or to send the instrument for in-house repair.


4. Further recommendations

This error can be avoided by ensuring proper connections between the attached bottles and the instrument prior to starting a program.


5. Next steps

• Note your instrument serial number.
• If the instrument is controlled by HydroControl™, save the log files.
• Create a service request and include the serial number and log files.

Tecan Helpdesk

 