Skip to content
  • There are no suggestions because the search field is empty.

ERR044 General information

1. Usual root causes
Unspecific hardware error of reader module.


2. First aid
• Disconnect the reader from the i-control™ or Magellan™ software.
• Switch the reader off and then on again.
• Connect the reader with i-control™ or Magellan™.

3. Typical solution
Service intervention is required to solve a hardware defect with the Infinite® 200 PRO.


4. Further recommendations
Create a service request and contact Tecan®.


5. Next steps
• Note the instrument serial number.
• Download and save log files.
• Create a service request and include the serial number and log files.

Tecan Helpdesk

 

Further Help

Infinite® 200 PRO How To 

 

Link to Tecan Product Page:

Infinite® 200 PRO

 

401817-011