- Tecan Knowledge Portal
- Microplate Reader
- Sunrise™ Troubleshooting
-
Liquid Handling
- Fluent® - How To...
- Fluent® - Troubleshooting
- Fluent® - Tips & Tricks
- Freedom EVO® - How To...
- Freedom EVO® - Troubleshooting
- Freedom EVO® - Tips & Tricks
- Freedom EVOlyzer® - How To...
- Freedom EVOlyzer® - Troubleshooting
- Freedom EVOlyzer® - Tips & Tricks
- EVO Logic™ - How To...
- EVO Logic™ - Troubleshooting
- EVO Logic™ - Tips & Tricks
- D300e Digital Dispenser - Tips & Tricks
- D300e Digital Dispenser - How To...
- Uno Single Cell Dispenser™ - How to...
- Uno Single Cell Dispenser™ - Tips & Tricks
- Resolvex® Prep - How To
-
Microplate Reader
- Infinite® F50 How To
- Infinite® F50 Trouble Shooting
- Infinite® 200 Pro How To
- Infinite® 200Pro Tips & Tricks
- Infinite® 200Pro Troubleshooting
- Sunrise™ How To
- Sunrise™ Troubleshooting
- Spark® How To
- Spark® Tips & Tricks
- Spark® Troubleshooting
- Magellan™ How To
- Magellan™ Tips & Tricks
- Magellan™ Troubleshooting
- MultiCheck™ How To
- Applications and Settings
-
Microplate Washer
-
Digital Solutions
-
General Information
-
Manuals
ERROR_0403 No Filter carriage 0403 detected
This issue requires a parts replacement via a service intervention
1. Usual root causes
The appearance of this error indicates a problem with the filter slide detection or the filter guide.
2. First aid
• Remove the filter slide from the instrument.
• Reinsert the filter slide into the instrument and check if error is resolved.
3. Typical solution
If the suggested actions do not resolve the error, the Sunrise™ hardware has a defect that requires a service intervention.
4. Further recommendations
Please create a service request and contact the Tecan® Helpdesk.
5. Next steps
• Note your instrument serial number.
• Create a service request and include the serial number.