- Tecan Knowledge Portal
- Microplate Reader
- Infinite® 200Pro Troubleshooting
-
Liquid Handling
- Fluent® - How To...
- Fluent® - Troubleshooting
- Fluent® - Tips & Tricks
- Freedom EVO® - How To...
- Freedom EVO® - Troubleshooting
- Freedom EVO® - Tips & Tricks
- Freedom EVOlyzer® - How To...
- Freedom EVOlyzer® - Troubleshooting
- Freedom EVOlyzer® - Tips & Tricks
- EVO Logic™ - How To...
- EVO Logic™ - Troubleshooting
- EVO Logic™ - Tips & Tricks
- D300e Digital Dispenser - Tips & Tricks
- D300e Digital Dispenser - How To...
- Uno Single Cell Dispenser™ - How to...
- Uno Single Cell Dispenser™ - Tips & Tricks
-
Microplate Reader
- Infinite® F50 How To
- Infinite® F50 Trouble Shooting
- Infinite® 200 Pro How To
- Infinite® 200Pro Tips & Tricks
- Infinite® 200Pro Troubleshooting
- Sunrise™ How To
- Sunrise™ Troubleshooting
- Spark® How To
- Spark® Tips & Tricks
- Spark® Troubleshooting
- Magellan™ How To
- Magellan™ Tips & Tricks
- Magellan™ Troubleshooting
- MultiCheck™ How To
- Applications and Settings
-
Microplate Washer
-
Digital Solutions
-
General Information
-
Manuals
Number must be either non-negative and less than or equal to Int32.MaxValue or -1. Parameter name timeout
1. Usual root causes
The error message “Number must be either non-negative and less than or equal to Int32.MaxValue or -1. Parameter name timeout” can appear during measurement with i-control™ or Magellan™ if Windows® updates are not performed on the computer controlling the instrument.
2. First aid
• Perform Windows® updates on the computer controlling the instrument.
• Install i-control™ V1.12 or higher for Infinite® 200 PRO, Infinite® M1000 PRO, Infinite® F500 and Infinite® F50.
• Install Magellan™ 7.2SP1 or higher for Infinite® 200 PRO, Infinite® M1000 PRO, Infinite® F500, Infinite® F50 and Sunrise™.
3. Typical solution
Perform Windows® updates as scheduled.
4. Further recommendations
If the issue cannot be resolved, create a service request.
5. Next steps
• Note the instrument serial number.
• Download and save log files.
• Create a service request and include the serial number and log files.