Using Daily Maintenance on Freedom EVOware®

How to perform Daily Maintenance on a Freedom EVO® instrument

This ticket describes how to run a Daily Maintenance script that can be imported as a template from the Freedom EVOware® Master CD.

For the full details about the Daily Maintenance for the Freedom EVO® instrument, please refer to the Operating Manual Doc 392886.

To enable the implementation of the Daily Maintenance routine, there is a template script available that can be imported into Freedom EVOware® Software and It can be adjusted to meet your specific instrument configuration. 

 

Maintenance for Liquid LiHa

1. Disposable Tips

Daily_EVO_01-1

2. Fixed tips

Daily_EVO_02-1

3. Liquid containers

Daily_EVO_03-1

4. Syringes and Tubing and Flushing the System

Daily_EVO_04-1

5. Worktable

Daily_EVO_05-1

6. Replacement/Removal of DiTi Adapter

Daily_EVO_06-1

 

Maintenance Air LiHa

1. Disposable Tip

Daily_EVO_01-1

2. Worktable

Daily_EVO_05-1

3. Replacement of the Air LiHa Inline Filter 

Daily_EVO_07-1

Further Recommendations

  • Fill the system liquid containers the night before to allow liquid to degas overnight.
  • To prevent deposits of crystals and growth of micro-organisms in liquid containers, clean all liquid containers at least once a week.
  • Make sure to allow solvents (e.g. ethanol) to fully evaporate before refilling reagents into the containers again.
  • Empty and Clean the waste container on the worktable at least once a day.
  • Perform the Daily Maintenance at the beginning and end of each day

 