What is the maximum volume that I can load into one Dispensehead of the cassette?

1. Maximum volume

A maximum volume of 20 µl can be dispensed from each Dispensehead of a T8+ cassette. The maximum volume decreases to 10 µl if at least one dispensed volume is below 1 nl.
If you want to dispense 20 µl of fluid with a single Dispensehead, please check your protocol and remove any dispensed volumes below 1 nl.

The maximum volume dispensable from a D4+ cassette is 200 µl. A D4+ cassette can only be used to dispense a fluid when there are no volumes below 1 nl specified in the protocol. D300eControl will automatically choose a T8+ cassette if at least one volume is below 1 nl.

Hover the cursor over the individual wells to check the volumes dispensed into each well and for the corresponding fluid (see image below, line highlighted by blue arrow).

 

KP_D300e_Q8

 

2. Further information

You can find further information about what can be dispensed with the D300e in the ‘Dispensing software user guide’, which that can be accessed from the D300eControl software through File > Documentation > Dispensing Software User Guide > Appendix E.

 

3. Further help

For further help, please refer to our Helpdesk.