How to remove the transport locks

The Infinite® 200 PRO is delivered with the plate carrier locked into place, so that it cannot be damaged. Before the instrument can be used, the transport lock must be removed:


1. Ensure that the instrument is disconnected from the main power supply.
2. Open the plate carrier compartment flap.
3. Remove the screws and pull the plate carrier out manually.
T86_1
4. Remove the screws from the transport lock.
T86_2
5. Remove the transport lock from the plate carrier.
T86_3
6. The transport lock should be saved for future transportation.

7. To reinstall the transport lock, perform steps above in reverse order.